Appendix A. HygroPro
II
HART
®
Field Device Specification
62
HygroPro
II
User's Manual
A.9.22 Command 161 (0xA1): Read Saturation Table Point
This command is to read a given temperature saturation table point.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.23 Command 162 (0xA2): Set Temperature Coefficients
This command sets the temperature coefficients.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned-8
Index of Saturation point.
Byte
Format
Description
0
Unsigned-8
Index of Saturation Table point.
1 ~ 4
Float
Tempearture deg C
5 ~ 8
Float
Saturation Constant
Code
Class
Description
0
Success
No Command-Specific Errors
1
Error
Index Out of Bounds.
2 ~ 127
Undefined
Byte
Format
Description
0 ~ 3
Float
Calibration Temperature
4 ~ 7
Float
Temperature Dependence
8 ~ 11
Float
Dampen Factor
Byte
Format
Description
0 ~ 3
Float
Calibration Temperature
4 ~ 7
Float
Temperature Dependence
8 ~ 11
Float
Dampen Factor
Code
Class
Description
0
Success
No Command-Specific Errors
1-4
Undefined
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......