DigitalFlow™ GF868 Service Manual (1 and 2-Channel)
3
Chapter 1. Calibration
1.3
Calibrating the Analog Outputs (cont.)
Figure 2: Ammeter Connection for Slots 1-6 (Output A)
1.3.1
Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate
the desired slot
. (The option bar will include a slot listing for each installed option card.)
3.
Press
[Fx]
to select the desired output.
IMPORTANT:
The procedure for calibrating
all the outputs is the same
. However, when calibrating
a different output
,
be sure to reconnect the ammeter to the appropriate pins on terminal block
I/O
. See Figure 1 on page 2
or Figure 2 above for the correct pin numbers.
Terminal Block (Option Card)
A
SIG
A
RTN
B
SIG
B
RTN
D
SIG
D
RTN
C
SIG
C
RTN
+
-
Ammeter
1
8