3.2 Displaying Diagnostic Parameters (cont.)
Table 3 lists the available diagnostic parameters for the Model GF868. The first column in the table shows the
parameter as it appears on the option bar, while the second column shows the parameter as it appears in the
prompt area after it has been selected.
Ch1 label gSITE
Press the [
and [
keys and the appropriate
[Fx]
key to display a different
diagnostic parameter or to
EXIT
the
Diagnostics Menu.
Ch1 DN Transit usec
519.7
(error codes appear here)
Tdow
n
DELT
A
PEAK
%
Qup
Table 3: Available Diagnostic Parameters
Option Bar
Screen Display
Description
Good
Bad
SS up
UP Sig Strength
Displays signal strength for the upstream transducer.
50–75
<50 or >75
SS do
DN Sig Strength
Displays signal strength for the downstream
transducer.
50–75
<50 or >75
SNDSP
Soundspeed Ft/s Displays the measured sound speed of the fluid.
N.A.
N.A.
Tup
UP Transit S usec Displays the upstream transit time of the ultrasonic
signal in microseconds.
N.A.
N.A.
Tdown
DN Transit usec
Displays the downstream transit time of the ultrasonic
signal in microseconds.
N.A.
N.A.
DELTA
DeltaT usec
Displays the difference in µsec between upstream and
downstream transit times.
N.A.
>10,000 nsec
Tot K K(Re)
MultiK*KFactor
Displays combination of all K factors.
N.A.
N.A.
PEAK%
PEAK%
Displays the percentage of peak (set to +50 by
default).
N.A.
N.A.
Qup
UP Signal Q
Displays the signal quality for the upstream
transducer.
1200
–400 to +400
Qdown
DN Signal Q
Displays signal quality for the downstream transducer.
1200
–400 to +400
AMPup
UP Amp Discrim
Displays the value for the amplitude discriminator of
the upstream transducer.
24 ± 5
<19 or >29
AMPdn
DN Amp Discrim Displays the value for the amplitude discriminator of
the downstream transducer.
24 ± 5
<19 or >29
CNTup
UP DAC Counts
Displays AGC DAC count for upstream gain setting.
N.A.
N.A.
CNTdn
DN DAC Counts
Displays AGC DAC count for downstream gain setting
N.A.
N.A.
P#up
UP +- Peak
Displays signal peaks for the upstream transducer.
100 to
2300
<100 or >2300
P#dn
DN +- Peak
Displays signal peaks for the downstream transducer.
100 to
2300
<100 or >2300
TEMP
Temperature F°
Displays the gas temperature (4-20 mA input).
N.A.
N.A.
PRES
Pressure psia
Displays the gas pressure (4-20 mA input).
N.A.
N.A.
Mw
Molecular Wt.
Displays the average molecular weight of the gas.
N.A.
N.A.
AcVOL
Act Vol. KACF/MIN Displays actual volumetric flow.
N.A.
N.A.
StVOL
Std Vol. KSCF/MIN Displays standard volumetric flow, if
STANDARD
equation is used.
N.A.
N.A.
Summary of Contents for GF868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 24: ...Chapter 1 Calibration DigitalFlow GF868 Service Manual 1 and 2 Channel 16...
Page 62: ...Warranty 60 DigitalFlow GF868 Service Manual 1 and 2 Channel...
Page 66: ...64 DigitalFlow GF868 Service Manual 1 and 2 Channel no content intended for this page...
Page 67: ......