DigitalFlow™ GS868 Programming Manual (2-Channel)
9
Chapter 1. Programming Site Data
2.
Press
[F1]
(OFF) to deactivate the channel and return to the
Channel PROGRAM
prompt, or press
[F2]
to activate
the channel in
BURST
mode.
3.
Press
[F1]
to select
Skan mode or
[F2]
to select
Skan/Measure mode. The meter will exit the
ACTIV
submenu and
return to the channel menu screen.
:As indicated in the above prompt, the Model GS868 flowmeter can take measurements in two different ways:
•
Skan
is a low resolution technique for locating the acoustic signal and for high velocity measurements. It is more
robust in a noisy environment than the Measure technique.
•
Measure
is a more precise technique best used for low velocity measurements.
If
Skan is selected at the next prompt, the instrument uses this technique exclusively. However, if S/M is selected, the
meter uses
Skan to find the acoustic signal and then tries to use the Measure technique for a more precise
measurement.
Note:
To change the
Skan
and
Measure
parameters, see the
SIGNL submenu
section on page 16.
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......