Chapter 1. Programming Site Data
8
DigitalFlow™ GS868 Programming Manual (2-Channel)
Programming of the
ACTIV
,
SYSTM
(
CH1
and
GLOBL
) and
PIPE
submenus is necessary for operation of the Model
GS868. Failure to accurately enter all of the required information will result in unreliable flow rate data. Therefore, be
sure to complete at least the sections of this chapter pertaining to those three submenus.
Note:
Because it is so essential, instructions for programming the
ACTIV
,
SYSTM
(
CH1
and
GLOBL
)
and
PIPE
submenus are also included in the Startup Guide. If that programming has already been completed, skip the
corresponding sections of this chapter.
Except for the three submenus noted above, it is not necessary to program the Model GS868 flowmeter in any
particular order. Therefore, the sections of this chapter need not be completed in sequence. Proceed immediately to
any section of interest.
1.6 Entering Programming Mode
To access the
User Program, press the
[PROG]
key on the keypad. The standard measurement mode screen will be
replaced by the following initial programming mode screen:
Note:
If the security feature is active, enter the password and press the
[ENT]
key to enter the User Program. See the
SECUR
submenu section on page 33 of this chapter for more information on the security feature.
1.7
Entering Channel Data
If
[F1]=CH1
(or
[F2]=CH2
) was selected at the initial
User Program prompt shown on page 1-8, the following screen
appears:
Based on the selection made above, proceed to the appropriate section of this chapter for instructions.
1.7.1
Activating a Channel
The
ACTIV
submenu activates a channel and selects the desired measurement method. While following the
programming instructions, refer to the menu map in Figure 11 on page 83
.
1.
Enter the
ACTIV
submenu by pressing
[F1]
at the
Channel PROGRAM
prompt.
PROGRAM Start
Press the
]
and
] k
eys and one of the function keys
[F1]-[F4]
to
select the desired submenu. From this screen, press
[EXIT]
to return to
measurement mode.
PROGRAM
Channel 1
CH1
CH2
GLOBL
SAVE
Chan 1 PROGR Start
Press the
]
and
] keys and a function key
[F1]-[F4]
to select the
desired submenu. From this screen, press
[EXIT]
to return to the initial
User Program screen shown on page 7.
Channel PROGRAM
status
ACTIV
SYSTM
PIPE
I/O
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......