
CGA 351 User’s Manual
69
Chapter 5.
Chapter 5.
5.1 Service and Maintenance
5.1.1
Introduction
The CGA 351 has been designed to provide years of trouble-free operation. However, because of the technical
challenges involved in measuring very low levels of oxygen, some difficulties may occasionally be encountered. The
procedures for resolving many of these situations are discussed in this chapter. If the information provided in this
chapter is insufficient to solve the problem, please contact the factory for further assistance.
If a faulty component is discovered in the CGA 351, that component is usually replaceable on site. Contact the factory
to obtain a new part, and proceed to the appropriate section of this chapter for installation instructions.
WARNING!
To ensure safe operation of the CGA 351, it must be installed and operated as described in this
manual.
In addition, be sure to follow all applicable local safety codes and regulations for installing electrical
equipment. In addition, all procedures should be performed by
trained service personnel.
5.1.2
General Troubleshooting
This section discusses the following general categories of problems that may arise with the CGA 351:
•
faulty wiring
•
a faulty thermocouple
•
a faulty oxygen sensor
•
a faulty sensor furnace
• missing sensor furnace voltage
Use this section to troubleshoot any malfunction in the CGA 351. Refer to Figure 5 on page 87, and complete the
following sections in the sequence indicated.
Note:
For information regarding component locations and/or wiring connections in an optional enclosure, see
Appendix C.
Summary of Contents for CGA 351
Page 2: ......
Page 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Page 32: ...CGA 351 User s Manual 26...
Page 94: ...CGA 351 User s Manual 88...
Page 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Page 104: ...Appendix B 98 CGA 351 User s Manual...
Page 112: ...CGA 351 User s Manual 106...
Page 118: ......