Chapter 4.
66
CGA 351 User’s Manual
Adjusting the Calibration Setting (cont.)
Note:
If the calibration “Failed”, repeat the calibration procedure. If another failure occurs, contact the factory for
assistance.
4.
Discontinue the flow of calibration gas to the analyzer.
5.
Resume the flow of process gas to the analyzer.
6.
If necessary, reconfigure the digital display to show the desired process parameter.
The CGA 351 may now be placed back into service.
4.1.15 Checking the Current Settings
To aid in monitoring the operation of the CGA 351, a built-in diagnostics routine is included in the
User Program. To
access this submenu, proceed as follows:
Table 122:
PPM-Percent..Cal-Gas
When ready, use the [
] and
[
] keys to select
[Next]
and
press
[ENTER]
.
Start the
Calibration Gas
Back
Next
Cancel
POWER
ESC
ENTER
Table 123:
PPM-Per..Cal-Heater
This message appears until
the calibration adjustment
has been completed (usually
about 15–30 minutes).
Calibrating Heater
S=xxx T=xxx.x D=xx.x
Back
Cancel
POWER
ESC
ENTER
Table 124:
If the calibration passed, use
the [
] and [
] keys to select
[Finish]
and press
[ENTER]
.
Calibrating Complete
Select Finish
Back
Finish
POWER
ESC
ENTER
Table 125:
1 Oxygen
The value shown is now the
correct ppm value for the
oxygen content of the
calibration gas.
5.00
ppm
POWER
ESC
ENTER
Summary of Contents for CGA 351
Page 2: ......
Page 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Page 32: ...CGA 351 User s Manual 26...
Page 94: ...CGA 351 User s Manual 88...
Page 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Page 104: ...Appendix B 98 CGA 351 User s Manual...
Page 112: ...CGA 351 User s Manual 106...
Page 118: ......