Aurora H2O User’s Manual
7
Appendix A. MODBUS RTU/TCP Communications
Appendix A. MODBUS RTU/TCP Communications
A.1 Introduction
The
Aurora H2O
supports digital communications using the Modbus/RTU protocol, with 2-wire RS-485 or 3-wire
RS-232C as the physical layer. Data rate can be specified at rates from 4800 to 115200 bits per second (bps), with
selectable parity.
The
Aurora H2O
supports the Modbus/TCP protocol as defined in: MODBUS Messaging on TCP/IP Implementation
Guide V1.0b.
Aurora H2O
has two physically separate communications ports. Both ports can be selected for either RS-232 or
RS-485 operation, and
Aurora H2O
can communicate with both ports simultaneously.
Note: Aurora H2O
supports the Modbus/RTU protocol as defined in:
MODBUS Application Protocol Specification, V1.1b
&
MODBUS over Serial Line Specification and Implementation Guide V1.02.
Note:
The above referenced specifications are available from the
Modbus Organization
at
http://modbus-ida.org.
A.2 Modbus Functions and Data Types
The Modbus functions supported by
Aurora H2O
include:
•
(0x03) Read Holding Registers
•
(0x04) Read Input Registers
•
(0x08) Diagnostics (Serial Line only) - only supports Echo subcommand
•
(0x10) Write Multiple registers
•
(0x11) Report Slave ID (Serial Line only)
•
(0x2B/0x0E) Read Device Identification - only supports Basic Device Identification tags, which are:
• VendorName
• Product code
• Revision number
Aurora H2O
supports data types of Integer and Double/Float. Integer data types are always four (4) bytes long and
should be read with a request for two registers (two bytes per each register, two registers total) at the address.
Double/Float data types provide eight (8) byte double precision data or four (4) byte single precision data,
depending on how many registers are requested. Four registers are used for double data types, and two registers are
used for single precision data types.
Aurora H2O Modbus register addresses are
zero-based
. Certain Modbus master software and devices use
one-based
addressing. With such devices, simply add 1 to the register addresses in
Table 7 on page 12
for correct
operation.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......