7
Maintenance
92
5200-en_V2.0_06/20
7.2
Replacing filters for pumps
The filter installed in the front plate of the controller protects the aerosol
pumps from dirt. A polluted filter causes the pumps to consume more en-
ergy. Replace the filter at least every 12 months or when the power con-
sumption of the aerosol pumps rises above 50%.
WARNING
Harmful substances in the filter pose a health risk.
If the cover on the filter is open, harmful substances from the filter can es-
cape into the ambient air and endanger persons.
4
Do not use compressed air when the filter cover is open.
4
Close the removed filter airtight and dispose of it properly.
4
Do not attempt to clean the filter. The filter cannot be regenerated.
Proceed as follows:
1.
Remove the clear protective cap from the filter by turning it to the left
and pulling it off.
If the cap cannot be turned by hand, use a pliers with soft contact sur-
faces. A cloth or a rubber mat can be used as contact surfaces.
2.
Check the state of the interior sealing ring (Ø 30 mm). Replace the
sealing ring if necessary.
3.
Extract the filter by turning it to the left and pulling it off.
4.
Insert a new or cleaned filter by turning it to the right.
5.
Turn the protective cap to the right by hand to put it back into place.