Nerve stimulation

MultiStim ECO

Summary of Contents for MultiStim ECO

Page 1: ...Nerve stimulation MultiStim ECO...

Page 2: ...other products are taken into account A decision regarding the combined use of devices from different manufacturers where they do not constitute treatment units is the responsibility of the user The d...

Page 3: ...e 93 42 EEC EN 62366 2008 EN 60529 2014 ISO 15233 1 2012 Intended use The MultiStim ECO is used to identify and locate peripheral nerves and nerve cords in local and regional anaesthesia for example d...

Page 4: ...as well as the applicable sub standards Please note that in the worst case all leakage currents and or the patient s aux iliary currents might add up so that inadmissibly high values put the patient...

Page 5: ...NK accessories Dynamic electric and magnetic interference fields might cause device system interactions They can affect the measurement of the actual stimulation current and in the worst case lead to...

Page 6: ...n rhythm can be heard i e stimula tion current flows through the patient If the green LED next to the stimulation button is on either the patient circuit is not closed or there is a resistance 12 k in...

Page 7: ...mulation current intensity 0 5 mA green LED below 0 5 mA is on Volume medium central green LED above the loudspeaker symbol is on Stimulation OFF LED next to stimulation button is off Function keys Bu...

Page 8: ...e device under any circumstances Send the Multi Stim back to the manufacturer Operation Check and start up Please note Devices showing other reactions must not be put into operation Please contact the...

Page 9: ...If this is the case you can start using it 5 As soon as you touch the skin surface with the cannula and or puncture the MultiStim ECO starts delivering current pulses You can perceive this visually by...

Page 10: ...xt to the unequal sign is also on As soon as the circuit is closed and pulses are delivered to the patient the yellow goes out At the same time the green stimulation LED starts blinking and the acoust...

Page 11: ...ly Only use alkaline manganese batteries e g Varta LR1 4001 LADY N Duracell LR1 LADY N Energizer LR1 LADY N Each time after replacing the batteries check the MultiStim ECO before start up to make sure...

Page 12: ...CIE Laboratoire Central des Industries Electroniques Aire de la Thur F 68840 Pulversheim Guidelinesandmanufacturer sdeclaration electromagneticinterference acc to EN 60601 1 2 2007 5 2 2 1 table 1 The...

Page 13: ...le 2 The MultiStim ECO is intended for use in electromagnetic environment as spe cified below Customers or users of the MultiStim ECO must ensure that it is operated in such an environment Immunity te...

Page 14: ...rt inter ruptions and variations of supply voltage according to IEC 61000 4 11 5 Ur 95 dip of Ur for period 40 Ur 60 dip of Ur for 5 periods 70 Ur 30 dip of Ur for 25 periods 5 Ur 95 dip of Ur for 5 s...

Page 15: ...3 Veffective value 150 kHz to 80 MHz within ISM bandsa Not appli cable Portable and mobile radio devices should not be used closer to the MultiStim ECO and its cables than the recommended safety dis t...

Page 16: ...ty that mobile portable communication devices cause interference if they are unintentionally brought into the patient s area For this reason an additional factor of 10 3 is used for calculating the re...

Page 17: ...g Maximum Output Charge per pulse Q2 0 mA I t Q2 0 mA 2 0 mA 100 s Q2 0 mA 0 2 As Maximum Average Current across a 500 ohm resistive load and at the maxi mum output setting 2 0 mA Iaverage I tw TStim...

Page 18: ...GmbH Medizintechnologie Karl Hall Strasse 1 78187 Geisingen Germany Key to symbols used in labelling Manufacturer Consult instructions for use Catalogue number Rx only Dispensing with prescription on...

Page 19: ...19 PAJUNK GmbH Medizintechnologie Karl Hall Strasse 1 78187 Geisingen Germany Phone 49 0 7704 9291 0 Fax 49 0 7704 9291 600 www pajunk com XS190244B_Englisch 2020 01 21 0124...