SECTION 5
REMOTE OPERATION
90
5.7
REMOTE CONTROL EXAMPLES
The REN (remote enable) line on the bus interface must be kept true to maintain REMOTE CONTROL.
REMOTE CONTROL can only be initiated while the UPC is displaying a Metering display or while in the UPC
STATUS display, not while editing.
Be sure that the UPC GPIB device address is set to a known unique address within the GPIB system. The
factory default device address is 1. See (
) for configuration of the GPIB device address.
for setting up the SERIAL INTERFACE parameters if this option is used.
Optional keywords have been omitted throughout this section for clarity. The command strings may contain
spaces.
NOTE:
Un-address means to send the UNTALK and UNLISTEN messages. This is only applicable for the
GPIB, NOT the Serial option. With many programming languages that support GPIB communications,
listener and talker addressing and un-addressing is handled automatically and is transparent to the user.
Check your programming language or driver documentation for verification.
NOTE:
<eos> should be a LINE FEED (0Ahex, 10dec). See GPIB convention #16 (
) for more details.
CAUTION
ALL :SOURce PARAMETERS SHOULD BE SET
PRIOR TO ENABLING THE OUTPUT FOR THE FIRST TIME
AFTER POWER-ON OR AFTER RESETTING THE UNIT.
THE PARAMETERS MAY BE SET INDIVIDUALLY OR
BY USING :DEFine AND :EXECute OF A STORED PROGRAM