Single Phase Mode Setup
For Single-phase mode current Calibration, the load must have the appropriate size. See next
section for recommended load value by model and phase mode.
AC Line
Input
DMM
Hi Sense
Lo Sense
Use optional P/N 160086 Single Phase
Mode Shorting Connector
R Load
A B C N N N
Figure 11-3: Current Calibration Equipment Setup – 1 Phase Mode
11.4.1
Current Calibration Load Values
Full-scale current gain calibration is best performed at current levels that are around 90% of full
scale current. Since the ADF Series® has only a single voltage range, maximum RMS current is
available at 120Vrms L-N for each phase. For single-phase mode, the current is three times
higher. Thus, to obtain 90% of load current, resistor value that is sized to draw this current level
at around 120Vrms is required. Since the maximum current varies by model and phase mode,
the table below shows the recommended resistor values or resistive load bank settings.
Three & Two Phase Mode
Single Phase Mode
Model
Rating/phs
VA/W
Max.
Irms/phs
Calibration
Current
(90%)
Recommended
R Load (Ohm)
Max/
Irms
Calibration
Current
(90%)
Recommended
R Load (Ohm)
190ADF
9000
75.0
67.5
1.07
390ADF
3000
25.0
23
5.33
1150ADF
15000
125.0
113
1.07
3150ADF
5000
41.7
38
3.20
3300ADF
10000
83.3
75
1.60
3450ADF
15000
125.0
113
1.07
3600ADF
20000
166.7
150
0.80
Table 11-3: Calibration Load Values by Model and Phase Mode