Pace 8007-0540 Manual Download Page 1

 

 

 

 

 

 

 
 

XR 3700 Real-Time

 

X-Ray Inspection System

 

(PN 8007-0542, 120V) 
(PN 8007-0546, 230V) 

 

 

XR 3500 Real-Time

 

X-Ray Inspection System

 

(PN 8007-0540, 120V) 
(PN 8007-0544, 230V) 

 

 

This manual operation is for 

PACE’s XR 3500 50kv unit and 

PACE’s XR 3700 70kv unit.

 

Summary of Contents for 8007-0540

Page 1: ...X Ray Inspection System PN 8007 0542 120V PN 8007 0546 230V XR 3500 Real Time X Ray Inspection System PN 8007 0540 120V PN 8007 0544 230V This manual operation is for PACE s XR 3500 50kv unit and PACE s XR 3700 70kv unit ...

Page 2: ...ty Guidelines 6 XR 3700 Specifications 7 Features 7 System Operation 8 Set up 9 XR 3700 Initial Power up and Operational Test 10 XR 3700 Maintenance 11 XR 3700 Troubleshooting 12 Design Safety Standards and Safe Operating Practices 13 Service and Warranty 15 PACE contact information and trademarks 16 ...

Page 3: ... aware that the ionizing radiation x rays can constitute a distinct hazard if not employed in strict accordance with instructions provided in this manual for maximum operator safety Exposure to excessive quantities of radiation can be dangerous to your health Avoid not only direct radiation exposure but also to secondary or scattered radiation which occurs when an x ray beam strikes or has passed ...

Page 4: ...ve a full understanding of how it works Additionally they should be familiar with established radiation safety exposure practices sanctioned by the National Bureau of Standards Handbook X ray Protection HB93 pertaining to x ray protection Service personnel should read this manual and be familiar with its contents before attempting to adjust or repair this equipment 2 Radiation Safety Information F...

Page 5: ...nd constraints regarding the modification of x ray systems conforming to 21CFR mentioned above Typically an x ray system must be registered with the appropriate state agency by the physical possessor of the system regardless of the legal owner Frequently registration must occur prior to the x ray system being placed into service Regulations may require initial and periodic inspections by a governm...

Page 6: ...mount of radiation being emitted by the machine is less than 0 5mR hour We recommend performing a radiation survey Following initial installation before placing the x ray system into service Whenever the x ray system is relocated Whenever the x ray system receives a strong jolt e g being dropped ore than one inch or struck sufficiently hard to dent or deform the exterior cabinet Whenever a leaded ...

Page 7: ...m PCB size 760mm x unlimited mm 30 x unlimited Focal spot 0 2mm 0 008 Focal spot to image plane distance 124mm 4 875 Adjustments Live or Capture video signal Frame averaging up to 32000 frames Video Gain adjustment Spatial Resolution 20lp mmv X Ray Actuation Foot Pedal Opening Clearance 40mm 1 5 120VAC 19mm 0 75 230 VAC PCB Fixture Device Standard Small PCB Carrier Standard Dimensions 394mm H x 45...

Page 8: ...determine the presence and degree of interlayer shifts At the onset of either conventional or small hole drilling inspection is used to qualify hole to pad alignment In addition to enhancing product quality x ray inspection enables users to control costs by eliminating defective boards early in the production process It can also be used to quality control incoming boards from vendors or customers ...

Page 9: ...te bnc cable should be connected to the composite video input on the video monitor a bnc to RCA converter might be needed The power cord should be connected to a properly grounded 120 VAC power supply B The XR 3500 3700 used with the TF 1700 2700 or IR3000 BGA CSP Rework Center Place the XR 3500 3700 on a stable work platform of sufficient capacity to adequately support the XR 3500 3700 The work p...

Page 10: ...has been returned to the off position The power light should blinks when the key switch is on and the cover lock is disengaged Then when the cover lock is engaged the power light turns steady 2 Make sure that there is nothing inside the inspection chamber close the cover if open and engage the lock and depress and hold the foot pedal The X ray will not come on if the cover is not locked into posit...

Page 11: ...deo processor included with this system is preset to 8 frames averaging Averaging makes the x ray image less grainy and can be adjusted to improve image quality but with an increase in motion lag 8 Images may be captured electronically when the XR 3500 3700 is used with the TF 1700 2700 PC or IR 3000 PC and software Your system is now ready for use 9 XR 3500 3700 Maintenance The XR 3500 3700 syste...

Page 12: ...ot the system Power The power category covers all problems that include lack of power to a component or to the full system Power problems would include The system not powering up The x ray controller s red light does not turn on The system s green light does not function Do all components power up when the main power switch is turned on Do all power indicating lamps turn on Do all power outputs sh...

Page 13: ...or Devices and Radiological Health Branch of the FDA as an Analytical X ray System As registered the system has a number of safety features provided to minimize any x ray scatter reaching the operator A Safety The X Ray tube assembly is lead shielded with at least a inch space between collimator and image plane to minimize any x ray scatter The collimator insures that the x ray beam size at the im...

Page 14: ...pecific regulations regarding the monitoring of possible radiation leakage of industrial x ray cabinets are determined by the individual state or country There are a number of measures which can be taken to provide a means of cursory monitoring D Radiation survey meters Radiation survey meters detect the presence of ionizing radiation and display a value in units of mR hr milli Roentgens per hour ...

Page 15: ...option repair or replace any defective products at its facility or other location approved by it at no charge to user or provides parts without charge for installation by the user in the field at user s expense and risk User will be responsible for all costs of shipping equipment to Seller or other location for warranty service EXCEPT FOR THE REMEDY ABOVE DESCRIBED UNLESS OTHERWISE REQUIRED BY APP...

Page 16: ...corporated to obtain the latest specifications The following are trademarks and or service marks of PACE Incorporated MD USA INSTACAL FUMEFLO HEATWISE PACEWORLDWIDE PERMAGROUND POWERMODULE and TEMPWISE The following are registered trademarks and or service marks of PACE Incorporated Laurel Maryland U S A ARM EVAC FLO D SODR MINIWAVE PACE SENSATEMP SNAPVAC SODRTEK SODR X TRACTOR THERMOFLO THERMOJET...