Page 3
Cleaning the POD
d. The POD can be cleaned after each patient use by using a solution of luke-
warm water and “Dettol” or similar
Disinfectant Fluid (use Dettol solution Instructions on “Dettol” Container). Remove the Mask and Filter from the
POD. Hold the POD firmly in one hand and unscrew (anti-clockwise) the White Rear Cap from the POD. Remove
the Main Diaphragm. Unscrew and remove the Dump Valve Cap from the top of the POD. Hold the POD upside
down so that the Hose is at the top preventing the cleaning solution running down into the Hose. By using a small
and smooth Brush/Pipe cleaner or cloth, thoroughly clean out the POD, Diaphragm and Caps using the cleaning
fluid solution. Rinse thoroughly and leave to drip dry or dry with a clean dry cloth. Gently reassemble then function
test and leak test as stated in 6.3 and 7c respectively. A full inspection should be carried out at least annually by
qualified personnel only. For Servicing/Maintenance Information please contact the Oxylitre Service Department.
8. Technical Data and Performance (POD Only)
a.
Input Pressure:
400 kPa. (Operates: from 280 to 600 kPa’s).
b.
Inspiratory Cracking Pressure:
up to 0.25 kPa at 10 LPM and up to 1.5 kPa at 200 LPM.
c.
Expiratory Resistance:
up to 0.1 kPa at 12 LPM.
9. Additional/Replacement Parts (Description/Part No.)
a. Filter: PRU010 (Box of 50)
b. Scavenging Outlet Connector: PRU0-20
c. Case: POK050
d. Wall Mounting Bracket: PRU003/W
e. Adult Mask: 011-4 (Single patient use) f. Child Mask: 011-2 (Single patient use)
g. Adult Mask: 012-4 (Re-usable, Autoclaveable) h. Child Mask: 012-2 (Re-usable, Autoclaveable)
i. PI Bodok Seal: OX010 (Regulator)
k. BN “O” Ring: BS110 (Regulator)
Note: The hose material is phthalate free PVC which is antimicrobial or is Silicone which complies with USP
Class VI.
POD Series Demand Valve Unit
For use with Medical Oxygen
Operating & Safety Instructions
Manufactured by:
Oxylitre Limited
Morton House
Skerton Road
Old Trafford
Manchester
M16 0WJ
England
Tel:
Fax:
email:
(0)161 872 6322
(0)161 848 7914
[email protected]
2797
GMDN 17169
(Demand Valve)
Doc Ref:
Issue No:
Date:
Doc No: Doc-Op-4802
Issue No:
2.8
Date Issue:
3.8.16