Oxylitre POD Series Operating/Safety Instructions Download Page 1

Page 3 

 

 

Cleaning the POD 

d.  The POD can be cleaned after each patient use by using a solution of luke-

warm water and “Dettol” or similar 

Disinfectant Fluid (use Dettol solution Instructions on “Dettol” Container). Remove the Mask and Filter from the 
POD. Hold the POD firmly in one hand and unscrew (anti-clockwise) the White Rear Cap from the POD. Remove 
the Main Diaphragm. Unscrew and remove the Dump Valve Cap from the top of the POD. Hold the POD upside 
down so that the Hose is at the top preventing the cleaning solution running down into the Hose. By using a small 
and smooth Brush/Pipe cleaner or cloth, thoroughly clean out the POD, Diaphragm and Caps using the cleaning 
fluid solution. Rinse thoroughly and leave to drip dry or dry with a clean dry cloth. Gently reassemble then function 
test and leak test as stated in 6.3 and 7c respectively. A full inspection should be carried out at least annually by 
qualified personnel only. For Servicing/Maintenance Information please contact the Oxylitre Service Department. 

8. Technical Data and Performance (POD Only) 

a. 

Input Pressure:   

 

400 kPa. (Operates: from 280 to 600 kPa’s). 

b. 

Inspiratory Cracking Pressure: 

up to 0.25 kPa at 10 LPM and up to 1.5 kPa at 200 LPM. 

c. 

Expiratory Resistance: 

 

up to 0.1 kPa at 12 LPM.

 

9. Additional/Replacement Parts (Description/Part No.) 

a. Filter: PRU010 (Box of 50) 

                         b. Scavenging Outlet Connector: PRU0-20  

c. Case: POK050 

 

                         d. Wall Mounting Bracket: PRU003/W 

e. Adult Mask: 011-4 (Single patient use)                f. Child Mask: 011-2 (Single patient use) 

g. Adult Mask: 012-4 (Re-usable, Autoclaveable)   h. Child Mask: 012-2 (Re-usable, Autoclaveable)      

 i. PI Bodok Seal: OX010 (Regulator)                      

k.  BN “O” Ring: BS110 (Regulator)                   

                        

 

Note: The hose material is phthalate free PVC which is antimicrobial or is Silicone which complies with USP 
Class VI. 

 

 

 

 

 

 

 

POD Series Demand Valve Unit 

For use with Medical Oxygen 

 

Operating & Safety Instructions 

Manufactured by:

 

Oxylitre Limited 

Morton House 
Skerton Road 
Old Trafford 
Manchester 
M16 0WJ 
England

 

Tel: 
Fax: 
email: 

(0)161 872 6322 
(0)161 848 7914 
[email protected] 

 

 

                         

2797 

 

 

GMDN 17169

 (Demand Valve) 

Doc Ref: 
Issue No: 
Date: 

Doc No: Doc-Op-4802 
Issue No: 

2.8 

Date Issue: 

3.8.16 

 

 

 

Reviews: