Cypher SPM
User Guide
5. Tighten the screws.
l
Using a 1.7 mm flat tipped screwdriver, tighten all the
screws.
l
First tighten them in the pattern shown for the first
four. When all 8 screws are snug, go around once
more and tighten firmly while holding the tool only
with fingertips to prevent over-tightening.
Note:
Be careful not to slip and possibly puncture
the membrane with the screwdriver.
6. Your parts should now be as shown below.
7. Preform the diaphragm.
l
Gently press the sample stage as shown so the dia-
phragm pops through to the other side.
Page 83 of 243
10 Cell Body and Sample Stage Guide