Electrical
Stage surface is not grounded (floating) and sufficiently isolated to be safe for use with high voltage
applications.
Temperature range
0 ºC to 120 ºC as measured by a sensor embedded several several mm below the surface on which
the sample sits.
Cleaning
When mounted to a CES cell body, it can be filled with nearly any solvent compatible with FFKM
and Stainless steel and quartz. When disassembled, wipe with a solvent soaked swab or cloth, but
keep liquids away from the back of the device and the circuit board connector.
Liquid use
Safe for use with samples in liquid droplets.
10.7 Reminders for cell exchange and handling
In this section, we provide a list of pitfalls to avoid during use of the Cypher ES for imaging. It is
intended for experienced Cypher S users who have had only basic training on the Cypher ES, provid-
ing a quick reminder of important warnings before using the Cypher ES.
1. The fitting wrench should never be used to attach gas lines to the cell body.
Using the fitting wrench on the cell body will crack the cell body glass.
l
The fitting wrench (114.800) supplied in the accessory
kit should NEVER be used with the cell bodies (see
top right image).
l
The fitting wrench should only be used with the fit-
ting fixture (114.721), which is also supplied in the
accessory kit, or the manifold on the front of the
scanner. See the bottom right image.
Page 77 of 243
10 Cell Body and Sample Stage Guide