Engaging The Cantilever On The Sample
Prerequisite: The sample must be loaded and the cantilever positioned about 1 mm above the
sample.
Loading the cantilever and sample on a standard scanner on page 198
Loading the cantilever and sample on an environmental scanner on page 205
.
Before you can perform imaging, you must engage the cantilever on the sample surface. This task
comprises three parts:
1. Bringing the cantilever tip to the sample.
2. Tuning the cantilever and setting scan parameters.
3. Landing the tip.
24.1 Bringing the cantilever close to the sample
Mode Master:
l
The software should now be showing the mode master window.
l
If not, click s the Mode Master button at the bottom of the screen:
.
2.
⇒
⇒
Select Mode Master Mode Example:
l
The ModeMaster Panel appears when you first open up the MFP3D software. Select AR Modes
⇒
Topography
⇒
ACAirTopography
3.