13. Attach a valve to one of the two tubing ends, as shown:
12.3 Seal the cell
Read ahead through the instructions and collect all the required items to place inside the glovebox.
1. Inspect the cantilever holder.
l
Ideally the cantilever holder has a hole in the pos-
ition shown with the cartoon hand. This will allow for
a tool to access a screw in a later step.
l
If there is no hole here, you can proceed without
using the screw, but should consider contacting sup-
[email protected] to see if your holder can
be updated.
2. Place all the collected items, the cell body with tubing and syringe attached and the cantilever
holder into the glovebox via the vacuum load lock.
Evacuating the lock will not harm the cell since it has not yet been sealed.
3. With the sample mount in the position shown, place the sample.
4. Load the cantilever into the cantilever holder and set aside.