13
Version1.0 rev 18 Feb 2015
Rotr
4. Attach adapters to the Rotr housing as per Andor Assembly drawing ENI-00766 (For adapter
part numbers refer to the sales order BOM).
5. Install the power cable and data cable per the connector IDs on the back of the filter wheel
(Section 2.2).
4.1.2.
A
SSEMBLY
This product itself requires no assembly. Cables connections are covered in Section N.N
4.1.3.
P
OWER
C
ONNECTION AND
P
ROTECTIVE
E
ARTHING
•
See Section 5.1 Emergency Disconnection
•
Before connection, check that the mains power socket used can provide the power as specified;
•
Rotr Section 7.1 Power Supply
•
Always switch off power before connecting/ disconnecting cables from the product.
•
A mains power cable is provided with this product, but ensure that it satisfies local regulations
for safety.
•
Mains power is connected via the 3-pin IEC C14 connector on the rear panel of the product.
•
An integral part of protection against electric shock in the case of a fault is the protective earth
provided via the earth conductors in the mains cable. It is therefore vital that the earth system of
the building, and in particular the socket, is constructed properly to provide suitable protection
when needed.
•
Do not pull cables by the sheath. Use the connector body.
4.2. I
NSTALLING
C
ABLES
4.2.1.
P
OWER AND
S
IGNAL
C
ABLING
Refer to Sections 2.2 and 7.1
4.3. I
NSTALLING THE
S
OFTWARE
Refer to the Software Guide supplied with your software for details.