MAINTENANCE
6.3 a
nnuaL
e
LectricaL
s
afety
c
hecks
It is advisable to check the integrity of the insulation and protective earth of the product on an annual basis, e.g. U.K.
PAT testing.
6.4 c
hanging
the
d
ichroic
(m
osaic
i
nfinity
)
The Mosaic Infinity features a manual dichroic slider (see also section 5.7) enabling the user to switch between the
installed dichroic filters. If a new dichroic is to be exchanged or added please follow the instructions outlined below:
WARNINGS:
•
Turn off lasers before beginning this procedure
•
Take care handling optical components, following standard practices for good optical handling techniques
1. Ensure that lasers are powered off.
2. Remove the 4 M3x8mm screws on the access panel on the dichroic module (shown below).
3. Remove the 4 M3x12mm screws which secure the required dichroic mount to the dichroic slider. Take care so
that the screws do not fall down into the microscope as they may damage the microscope’s objective lens
4. Fasten the new dichroic into the required position.
5. Fasten the access panel to the dichroic module by fastening the 4 screws previously removed.