9.Technical Specifications
73
9.
Technical Specifications
Unless otherwise specified, the technical specifications applied are for TDS series
only, and Probes attenuation set as 10X. Only if the oscilloscope fulfills the following
two conditions at first, these specification standards can be reached.
This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
If change of the operating temperature is up to or exceeds 5
℃
, do a
"Self-calibration" procedure (see "
How to Implement Self-calibration
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Performance Characteristics
Instruction
Bandwidth
TDS7104
TDS8104
100MHz
TDS8204
200MHz
Channel
4
Acquisition
Mode
Normal, Peak detect, Averaging
Waveform Capture
rate
50,000
Max Sample rate
(real time)
TDS7104
Four CH
500MS/s
Dual CH
1GS/s*
Single CH
1GS/s
TDS8104
TDS8204
Four CH
1GS/s
Dual CH
2GS/s*
Single CH
2GS/s
Input
Input coupling
DC, AC , Ground
Input impedance
1M
Ω±2%, in parallel with 10pF±5pF
50Ω±1%
Probe attenuation
factor
1X, 10X, 100X, 1000X
Max. input voltage
1MΩ input impedance:
400V (PK-PK) (DC + AC PK-PK)
50Ωinput impedance:
5V (PK-PK) (DC + AC PK-PK)
Bandwidth limit
20MHz, full bandwidth
Channel
–channel
isolation
50Hz: 100 : 1
10MHz: 40 : 1