5
.Menu Interpretation
51
a user-defined idle time (1 to 60 minutes). Press the power key to
re-activate the LCD display after the LCD display sleeping mode
has been triggered.
5.10
【
Sweep
】
hardkey
Key access:
Press
Shift key, then
Sweep
hardkey (
4
in
numeric keypad
).
Sets parameters about the Sweep time and mode including
[Sweep Time Auto Manual]. [Sweep Single], [Sweep Cont], [Sweep
Points].
5.10.1 [Sweep Time Auto Man]
Sets the time interval for the analyzer to complete a sweep.
In non-zero span, the analyzer uses the shortest sweep time on
the basis of the current RBW and VBW settings if Auto is
selected.
You can modify this parameter using the numeric keys, knob, or
direction keys.
5.10.2 [Sweep Single]
Press to set the sweep mode to single sweep. In this mode,
pressing [Seep Single] enables a sweep.
5.10.3 [Sweep Cont]
Press [Sweep Cont] to set the sweep mode to continuous sweep.
The analyzer performs one sweep after another as soon as it is
triggered.
5.10.4 [Sweep Points]
Sets the desired points for every sweep. That is the number of
points of the current trace.