USER MANUAL
Error messages on the display
(Rev. 6)
Ow-RX
59
9.
ERROR MESSAGES ON THE DISPLAY
Ow-RX is fully controlled by a microprocessor which controls the
programming of the emission parameters and signals the various
conditions of the machine, the possible abnormalities and errors via
displayed messages.
The following tables show the various messages that can appear on the
display, their meaning and their cause.
The error messages are separated into three different categories,
classified based on the severity of the abnormality discovered and their
possible effect on the safety of the operators and/or the system.
9.1
Fatal errors upon power-up and in the ready, idle
and cooling statuses
These signals do NOT permit an examination to be performed.
It is possible to try to turn the equipment off and on, but if the signal is
repeated, technical assistance must be contacted.
Displayed
message
Type of ABNORMALITY
ACOUSTIC
signal
E01
X-ray button pressed at power-up
None
E02
A key pressed at power-up
(other than the X-ray button)
None
E03
Multiple keys pressed at power-up
None
E05
Unwanted X-ray emission
Present as long
as RX ON is
active
WARNING:
If E01 is displayed, release the X-ray button; if this has not been
pressed, this indicates a fault, therefore call the support service.
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...