User Manual
– Technical characteristics
NIMXEN020I
Owandy Radiology SAS
23
6.
TECHNICAL CHARACTERISTICS
General features
Type
I-MAX 3D
Manufacturer
Owandy Radiology
Class
Class I with type B applied parts
according to IEC 60601-1
classification.
Protection degree
IPX0 standard device
Line voltage
99-132V
198-264 V
Rated line voltage
110-120V
220-240V
Line frequency
50/60Hz
Maximum line current
14.5A @115V 50/60Hz
6A @ 230V 50/60Hz
Technical factors for maximum line current
86kV, 12.5mA
Power consumption
1.5kVA @ 115V 50/60Hz
1.3kVA @ 230V 50/60Hz
Protection fuse (F1)
20 A T 250V 6.3x32 mm 10kA@125V
8 A T 250V 6.3x32 mm 200A@250V
Column protection fuse (F2)
3 A T 250V 6.3x32 mm 10kA@125V
1.6 A T 250V 6.3x32 mm 100A@250V
Line apparent resistance
0.2
Ω max (99-132 V)
0.5
Ω max (198-264 V)
Line voltage regulation
< 3% @ 99 V~
Rated output voltage (kV)
60
– 86 kV, with 2 kV steps
Anodic current
2 - 12.5mA, with R20 scale steps (2, 2.2, 2.5,
2.8, 3.2, 3.6, 4, 4.5, 5, 5.6, 6.3, 7.1, 8, 9, 10, 11,
12.5)
Total filtration
≥ 2,5 mm Al eq. @ 86 kV
ref. IEC 60601-1-3 Par. 7.1