User Manual - Specification of the intended use
NIMXEN020I
Owandy Radiology SAS
4
2.3.1
Panoramic mode
The air kerma value at the entrance of the X-ray image receptor for the PANORAMIC exam is
reported in the table below as functions of kV and mA.
kV 60
62
64
66
68
70
72
74
76
78
80
82
84
86
mA
Air Kerma [mGy]
2
3.57
3.85
4.14
4.43
4.72
5.00
5.29
5.58
5.86
6.15
6.44
6.61
6.75
6.88
2.2
3.93
4.24
4.56
4.87
5.19
5.50
5.82
6.13
6.45
6.76
7.08
7.27
7.43
7.57
2.5
4.46
4.82
5.18
5.54
5.89
6.25
6.61
6.97
7.33
7.69
8.04
8.26
8.44
8.60
2.8
5.00
5.40
5.80
6.20
6.60
7.00
7.40
7.81
8.21
8.61
9.01
9.25
9.45
9.63
3.2
5.71
6.17
6.63
7.09
7.54
8.00
8.46
8.92
9.38
9.84
10.30 10.57 10.80 11.00
3.6
6.42
6.94
7.45
7.97
8.49
9.00
9.52
10.04 10.55 11.07 11.58 11.89 12.15 12.38
4
7.14
7.71
8.28
8.86
9.43
10.00 10.58 11.15 11.72 12.30 12.87 13.21 13.50 13.76
4.5
8.03
8.67
9.32
9.96
10.61 11.25 11.90 12.54 13.19 13.83 14.48 14.86 15.19 15.48
5
8.92
9.64
10.35 11.07 11.79 12.50 13.22 13.94 14.65 15.37 16.09 16.52 16.88 17.20
5.6
9.99
10.79 11.60 12.40 13.20 14.00 14.81 15.61 16.41 17.22 18.02 18.50 18.90 19.26
6.3
11.24 12.14 13.05 13.95 14.85 15.76 16.66 17.56 18.46 19.37 20.27 20.81 21.27 21.67
7.1
12.67 13.69 14.70 15.72 16.74 17.76 18.77 19.79 20.81 21.83 22.85 23.45 23.97 24.42
8
14.27 15.42 16.57 17.71 18.86 20.01 21.15 22.30 23.45 24.59 25.74 26.43 27.01 27.51
9
16.06 17.35 18.64 19.93 21.22 22.51 23.80 25.09 26.38 27.67 28.96 29.73 30.38 30.95
10
17.84 19.27 20.71 22.14 23.58 25.01 26.44 27.88 29.31 30.74 32.18 33.03 33.76 34.39
11
19.63 21.20 22.78 24.36 25.93 27.51 29.09 30.66 32.24 33.82 35.39 36.34 37.13 37.83
12.5
22.30 24.09 25.89 27.68 29.47 31.26 33.05 34.84 36.64 38.43 40.22 41.29 42.20 42.99