Error message
Error no.
Content
Cause
Possible measure
Defective joystick
Orientation Error
(None)
Orientation error on
advanced
stability
module (ASM)
Defective advanced
stability
module
(ASM)
Contact
specialist
dealer
ASM Disconnected
(None)
Advanced
stability
module (ASM) dis
connected from con
troller
e.g. defective
plug
connection,
cable
break
Defective advanced
stability
module
(ASM)
Contact
specialist
dealer
Gyro Disconnected
(None)
Electronic track sta
biliser (gyro module)
disconnected
from
controller
e.g. defective cable
connection to con
troller
Defective gyro mod
ule
Contact
specialist
dealer
Gyro Prf Active
(None)
Electronic track sta
biliser (gyro module)
is activated
Electronic track sta
biliser (gyro module)
is disrupted by the
movement of a differ
ent means of trans
portation (e.g. train,
ship) in standby
Defective gyro mod
ule
When driving/stand
ing in means of trans
portation, select the
"No Assist" driving
profile for safety reas
ons
If error persists: con
tact specialist dealer
1E0D
Short
circuit
on
external on/off switch
Switch Short
1E0C
Short
circuit
on
external profile/mode
switch
e.g. defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists: con
tact specialist dealer
7.4.4 Attendant control error overview
VR2 attendant control
Flashing LED
Error/warning
Cause
Possible measure
Control unit fault
Fault in the control unit
Read error on the control panel and
take
corresponding
action
(see
"Wheelchair control unit error over
view" table)
Attendant control fault
Defective attendant control Switch off attendant control and turn
control panel off/on
Joystick error
Attendant control joystick
not in home position when
the unit is turned on
Switch off attendant control and move
joystick to home position before
switching the unit on
Switch control panel off/on
Communication
error
between the attendant con
trol (joystick) and the con
trol panel/controller
Defective cable, loose plug
connection
Check cabling / plug connections
R-Net attendant control
Flashing LED
Error/warning
Cause
Possible measure
Control unit fault
Fault in the control unit
Read error on the control panel and
take
corresponding
action
(see
"Wheelchair control unit error over
view" table)
109
Juvo
Maintenance and repair
Summary of Contents for Juvo
Page 1: ...Juvo Instructions for use user 3...
Page 2: ...2 Juvo...
Page 120: ...Juvo 120...
Page 121: ...121 Juvo...
Page 122: ...Juvo 122...