100
16.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Event
Required action
–
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
Overheated hydraulic
unit
Reduce activity.
–
3 x long
Charge level under
25%
Charge battery soon.
Remaining operating time
approx. 24 hours
–
5 x long
Charge level under
10%
Charge battery soon
Remaining operating time
approx. 6 hours
5 x long
5 x long repeated
every 60 seconds
Medium error (see
Page 92)
e.g. a sensor is not
operational
Walking possible with
restrictions. Please note
the change in flexion res
istance.
The product must be
inspected immediately by
an O&P professional.
10 x long
10 x long
Battery charge level
5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Charge the battery.
30 x long
1 x long, 1 x short
repeated every
3 seconds
Severe error/indic
ation of safety
mode activation
(see Page 92)
e.g. one or more
sensors are not oper
ational
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected immediately by
an O&P professional.
–
Continuous
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected immediately by
an O&P professional.