10 Technische Daten
Kennzeichen
4R84
Gewicht [g]
115
Systemhöhe [mm]
36
Einbauhöhe [mm]
36
Material
Titan
Max. Körpergewicht [kg]
150
1 Product description
English
INFORMATION
Date of last update: 2022-03-07
►
Please read this document carefully before using the product
and observe the safety notices.
►
Instruct the user in the safe use of the product.
►
Please contact the manufacturer if you have questions about the
product or in case of problems.
►
Report each serious incident related to the product to the manu
facturer and to the relevant authority in your country. This is par
ticularly important when there is a decline in the health state.
►
Please keep this document for your records.
1.1 Construction and Function
The 4R84 double adapter features a pyramid adapter and pyramid
receiver. It serves as a connector between prosthetic components.
Proximal and distal angle adjustments are possible in the frontal and
sagittal planes.
1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of
modular connectors. Functionality with components of other manufac
turers that have compatible modular connectors has not been tested.
6
2 Intended use
2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fit
tings.
2.2 Area of application
Approved for a body weight of
up to 150 kg
.
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range:
–10 °C to +45 °C (14 °F to 113 °F)
Chemicals/liquids:
fresh water, salt water, perspiration, urine,
soapsuds, chlorine water
Moisture:
Submersion: max. 1 h in 3 m depth, relative humidity: no
restrictions
Solids:
Dust, occasional contact with sand
Clean the product after contact with humidity/chemic
als/solids, in order to avoid increased wear and damage (
see
page 8
Unallowable environmental conditions
Solids:
Highly hygroscopic particles (e.g. talcum), continuous con
tact with sand
Chemicals/liquids:
Acids, continuous use in liquid media
2.4 Lifetime
This product was tested by the manufacturer with 3 million load
cycles. Depending on the user’s activity level, this corresponds to a
maximum lifetime of 5 years.