![Otto Bock 17KO1 Series Instructions For Use Manual Download Page 101](http://html1.mh-extra.com/html/otto-bock/17ko1-series/17ko1-series_instructions-for-use-manual_1659386101.webp)
Switching process
1) On the screen of the device, tap the corresponding symbol to activate training mode.
→
A confirmation signal sounds to indicate the switch to training mode.
2) The hydraulics stabilise the joint with high flexion resistance in the stance phase and
release the joint in the swing phase so that the leg can swing forward freely.
Exit training mode by selecting a different MyMode or by turning the component off and then
on again.
11.4 MyMode "Freeze position"
(if the knee joint is permanently locked)
Switching process
In this MyMode, the joint is locked in its current position and cannot be moved in the flexion
nor the extension direction.
1) Flex or extend the knee joint to the angle where it should be locked.
2) On the screen of the device, tap the corresponding symbol to activate the lock.
→
A confirmation signal sounds to indicate activation of the lock.
3) The orthosis is locked in the flexion and extension directions.
Deactivate the lock by selecting a different MyMode/basic mode or by turning the compon
ent off and then on again.
11.5 MyMode "User defined"
The MyMode "
User defined
" is intended for specific motion patterns or postures
(e.g. cycling ...). This MyMode can be configured individually using the adjustment
app.
Settings can also be adjusted by the patient using the Cockpit app (see Page 97).
101
C-Brace joint unit 17KO1=*
MyModes
Summary of Contents for 17KO1 Series
Page 4: ...4 C Brace joint unit 17KO1...
Page 114: ...114 C Brace joint unit 17KO1...
Page 115: ...115 C Brace joint unit 17KO1...