Test device
Problem
Possible cause
Solution
The wrong test device was selec-
ted.
See
Test device does not perform
correctly
Battery status low in test device. Recharge the
test device.
Air hose/cable connection prob-
lems
Check the connections. See
17.4
Probe-fit related problems
Leakage
Problem
Possible cause
Solution
Leakage
Leakage can be caused by a num-
ber of factors, such as:
•
Pneumatic/probe plug not inserted prop-
erly.
•
Inappropriate ear tip size (either too small
or too large).
•
Ear canal debris blocks for good ear tip seal.
170
Otometrics - MADSEN OTOflex 100
17 Troubleshooting