Max (dB)
Select the desired maximum level (up to 110 dB).
Warning
•
Be careful when you enter the Max setting, so as
not to expose the patient to too high intensity levels!
You can extend the range to 105 dB (120 dB in Contra). If you do, a
warning will appear on the screen, and the test screen will visually
indicate that you are using extended range!
Start (dB)
Select the desired start level (from 40 to 110 dB).
Increment (dB)
Select the step size to increment the stimulus intensity (Sound Pres-
sure Level) in increments of 1, 2, 5, or 10 dB.
Default: 10 dB
Stimulus Timing
Pre-stim(ulus time)
Defines the recording time before each stimulus is presented.
Default value: 0.2 seconds
Stimulus (time)
Defines the duration of the stimulus. Displays the text
Manual
when
set to 0.
Default value: 1.3 seconds
Post-stim(ulus time)
Defines the recording time after each stimulus is terminated.
Default value: 0.5 seconds
Pause (time)
Defines the inter-stimulus interval.
Default value: 0.5 seconds
Stimulus types
(Auto-
matic)
Stimulus types
Click to select the stimulus types to be included/excluded. At least
one stimulus type must always be selected.
Available types: 500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN, LBN, and
HBN.
Verification
Verification
Select the type of verification strategy to be used after the deflection
criterion has been met by the automatic threshold search.
The stimulus may be retested for reproducibility, or the next one or
two intensity levels can be used to confirm the reflex growth prop-
erties.
Default:
Do next level
14.3.5
Reflex Decay setup
When you perform a Reflex Decay measurement, there are a number of settings you can adjust. These settings are avail-
able either in the Control Panel on the left side of the screen, or in the
Tools
>
Options
>
Reflex Decay
>
Measurement
dialog.
Otometrics - MADSEN OTOflex 100
157
14 Immittance module settings