Auto next curve
Menu
>
More settings..
Select
On
or
Off
to define whether the system will automatically select the next curve
number available for the new measurement results when you press
Start
to start a new
measurement.
Example:
If the setting is
Off
, and the curve number shown is no. 3, the results currently applying to
this number will be overwritten.
Default:
On
Press(ure) range
Menu
>
More settings..
The pressure range selected in the measurement setup.
You can choose between
Normal
and
Custom
pressure range.
•
Normal
The normal range of pump pressure is from -400 daPa to +200 daPa.
•
Custom
The custom range (Custom min. press. and Custom max. press. settings, listed below)
can be accessed if
Custom
is selected.
You can extend the pressure range of –400 to +200 daPa to as much as –600 to +400 daPa,
or decrease it to -50 to +50 daPa.
Default: Normal
Custom min. press(ure)
Menu
>
More settings..
Custom minimum pressure ranges from -50 to -600 daPa in steps of 50 daPa.
Default: -600 daPa
Custom max. press(ure)
Menu
>
More settings..
Custom maximum pressure ranges from +50 to +400 daPa in steps of 50 daPa.
Press
Select
and scroll to the appropriate pressure value.
Default: 400 daPa
Pump speed
Menu
>
More settings..
Changes the speed of the pressure sweep. Indicated in daPa per second.
•
50, 100, 200, 400 daPa/s, or
•
AFAP
(As Fast As Possible, 500-600 daPa/s in 2 - 0.5 cc): Forces the pump to work as
fast as possible. The actual pressure build-up depends on the amount of air escaping
from probe and ear.
Infant default:
AFAP
daPa/s
Adult default: 200 daPa/s
Otometrics - MADSEN OTOflex 100
133
13 Test settings in MADSEN OTOflex 100