![otometrics MADSEN OTOflex 100 Reference Manual Download Page 128](http://html.mh-extra.com/html/otometrics/madsen-otoflex-100/madsen-otoflex-100_reference-manual_1658679128.webp)
Intensity safety
On
:
When Intensity safety is on, no stimulus intensities higher than 115 dB SPL (measured in a 2cc
cavity) can be selected or will be applied.
Off
:
Stimulus intensities up to the technical limits of the device and probe can be selected and
applied. A warning triangle is displayed on-screen whenever intensities above 115 dB SPL are
selected.
Intensity safety is automatically set to
On
whenever the
Ear Selector
button is toggled or a
new patient folder is selected.
Note
•
Whenever a stimulus level exceeds the warning level (> 108 dB SPL re 2cc), the
stimulus intensity value will start flashing and all automatic testing is paused. You are
then prompted to decide whether to continue or to move on to the next stimulus type.
In Manual testing, whenever a stimulus level exceeds the Intensity Safetp.buy level
((> 115 dB SPL re 2cc), the stimulus intensity value will start flashing.
Warning
•
When you test on small ears, the sound pressure level will increase in the
ear canal. It is therefore not recommended to exceed the warning level when testing
on patients with small ear canals.
Default:
On
Quick flow settings
See
for a description of the settings below.
•
‘Done?’ prompt
•
Print when done
•
‘Patient’ prompt
•
‘Settings’ prompt
13.2.3
Advanced
These settings apply regardless of the test type or sequence selected. To access these settings:
•
Select
Menu
>
Advanced..
128
Otometrics - MADSEN OTOflex 100
13 Test settings in MADSEN OTOflex 100