Speech testing
To perform speech testing, see
•
Special tests
SISI (Short Increment Sensitivity Index)
,
ABLB (Alternate Binaural Loudness Balance)
Hearing Instrument Simulation (HIS)
These tests are optional. To add any of these tests to your configuration, contact your local supplier.
5.6
Proper transducer placement
Headphones
1. Loosen the headband and place both the left and right side of the headphones simultaneously.
2. Make sure that the blue indicator is on the left ear and the red indicator is on the right ear. The center of the head-
phone diaphragm should be directly opposite the opening of the ear canal for the left and right side.
Note
•
If the headphones are not placed properly, there is risk of causing the ear canal to collapse which will result in
elevated thresholds.
3. Ask the patient if the headphones are comfortable.
4. Tighten the headband while holding the headphones in place with your thumbs.
5. Examine the placement of the headphones to make sure they are level, and properly positioned.
Insert Earphones
Young children tolerate insert earphones better than headphones.
The insert earphones are color-coded:
•
the blue indicator for the left ear,
•
the red indicator for the right ear.
1. Select the largest foam eartip that will fit into the patient's ear.
If the eartip is too small the sound will leak out and the dB level will not be accurate at the eardrum.
Insert earphones have greater attenuation between ears especially at the low frequencies; this reduces the need for
masking.
2. It is best to clip the insert earphone transducers behind the child or on the back of their clothing and then fit the
foam eartip into the child's ears.
If the foam eartip is a little too large, consider cutting it down to make it a little smaller.
Bone Oscillator
Note
•
For unmasked bone thresholds, you can store binaural data:
58
Otometrics - MADSEN Itera II
5 Preparing for testing
Summary of Contents for madsen itera ii
Page 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Page 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Page 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Page 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Page 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Page 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Page 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Page 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Page 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...