4.9.1
The work area in the Speech screen
Selecting tabular or graphical view
In the Speech test screen of the OTOsuite Audiometry module you can use either tabular view or graphical view.
1. To select the view you wish to use, select
Tools
>
Options
>
Audiometry
>
Speech
>
Speech View Mode
>
Tabular
View
or
Graphical View
.
The work area can be shown in two modes:
•
•
4.9.2
Speech feature boxes
You can access a number of speech display options directly from the
Speech
main screen.
Feature boxes
Pure Tone Data
Displays the pure tone average for air conduction and bone conduction as well as the articulation
index.
•
To configure pure tone average (PTA) calculation, select
Tools
>
Options
>
Audiometry
>
Gen-
eral
>
Misc.
>
PTA Frequency AC/BC
.
•
The AI is calculated according to the “Count-the-dot” method.
Stenger
Displays the results of a speech Stenger test (optional).
•
Scoring options are positive (+) or negative (-).
4.9.3
Terms and abbreviations used in Speech testing
SDT
Speech Detection Threshold
SRT
Speech Recognition Threshold
MCL
Most Comfortable Loudness Level
UCL
UnComfortable Loudness Level
WRS/SRS
Word Recognition Score/Sentence Recognition Score
•
Score
Percentage of correct/incorrect words.
•
Level
dB level at which the words were presented.
•
[Msk]
Effective masking level (dB EML) used for contralateral masking.
•
SNR
Signal-to-Noise Ratio.
Otometrics - MADSEN Itera II
47
4 Navigating in the OTOsuiteAudiometry Module
Summary of Contents for madsen itera ii
Page 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Page 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Page 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Page 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Page 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Page 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Page 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Page 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Page 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...