Technical specifications
19.1
MADSEN Itera II
Type identification
MADSEN Itera II is type 1004 from Natus Medical Denmark ApS.
Channels
2 separate and identical channels
Pure tone frequencies
AC and SF:
12 standard 125 - 8000 and 12500 Hz
BC:
250 - 8000 Hz standard frequencies
Insert phones
125 - 6000 Hz standard frequencies
Accuracy:
Better than 1 %.
Modulation
FM (Warble):
1 - 20 Hz in 1 Hz steps. Mod. width 1% - 25% in 1% steps
AM for SISI:
5, 4, 3, 1, 0.75, 0.50, 0.25 dB HL steps
Wave form:
Triangular
Attenuator
1 dB HL / 2.5 dB HL / 5 dB HL step resolution over the entire range
Attenuator accuracy
In whole range:
better than 3 dB HL
Between two consecutive attenuator positions:
5 dB HL step:
better than 1 dB HL
2.5 dB HL step:
better than 0.75 dB HL
1 dB HL step:
better than 0.3 dB HL
HL Range
Maximum output will be limited by the transducer.
AC:
-10 to 120 dB HL at mid-frequencies
BC:
-10 to 70 dB HL at mid-frequencies
Otometrics - MADSEN Itera II
119
Summary of Contents for madsen itera ii
Page 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Page 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Page 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Page 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Page 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Page 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Page 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Page 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Page 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...