16
Maximum output levels
16.1
Tone
Frequency
TDH39
Bone conductor
(B-71)
Insert phone
(E-A-RTONE® 3A)
HDA 200
HolmCo
125
80
105
100
85
250
100
45
110
110
100
500
120
65
120
120
115
750
120
65
120
120
120
1000
120
75
120
120
120
1500
120
80
120
120
120
2000
120
80
120
120
120
3000
120
80
120
120
115
4000
120
75
120
120
115
6000
110
50
110
110
110
8000
100
45
100
105
110
9000
n/a
100
10000
n/a
90
11200
n/a
85
12500
75
80
14000
60
16000
50
n/a: data not available
16.2
NBN
Frequency
TDH39
Bone conductor
(B-71)
Insert phone
(E-A-RTONE® 3A)
HDA 200
HolmCo
125
65
95
80
50
250
95
n/a
105
90
100
500
115
n/a
110
100
110
750
115
n/a
110
100
115
1000
115
n/a
115
100
115
1500
115
n/a
115
95
115
Otometrics - MADSEN Itera II
111
Summary of Contents for madsen itera ii
Page 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Page 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Page 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Page 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Page 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Page 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Page 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Page 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Page 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...