Caution
•
Do not allow any moisture inside the instrument!
•
To clean the display, use a dry cloth or soft brush. Note that the display is laminated for maximum readability and must
therefore be treated with care.
Headphones
The headphones are in constant contact with the patient, and should therefore be kept clean.
Clean the headphones between patients, e.g. with a non-alcohol based antibacterial wipe, such as Audiowipes.
Eartips for Insert Earphones
Warning
•
To prevent cross-infection, use new eartips when you test the next client.
The eartips are disposable and therefore should not be cleaned or re-used. There are no special requirements for the dis-
posal of the eartips.
Bone oscillator
Clean the bone oscillator between patients, e.g. with a non-alcohol based antibacterial wipe, such as Audiowipes.
13.3
Calibration
You can download new software to MADSEN Itera II, and if this involves no change to the transducers, MADSEN Itera II
does not need recalibration.
Annual calibration
The audiometer, headphones, bone oscillators, and sound field speakers must be calibrated once a year by your authorized
service department.
The audiometer is dispatched from the factory together with a Test Report (Calibration Certificate). The Test Report spe-
cifies the transducers that have been calibrated (i.e., those which have been supplied together with the instrument),
according to which standards, and the equipment used for calibration. Results are listed for each transducer at all standard
frequencies.
In general, the instrument is calibrated in dB HL and dB masking level using the stated reference equivalent thresholds; dB
HL is related to sound pressure levels, dB SPL = dB re 20
µ
Pa, and force levels (dB re 1
µ
N).
Caution
•
Note that calibration has been performed only on the transducers supplied! If you wish to use any other
transducer for testing with the device, please contact your local distributor first.
104
Otometrics - MADSEN Itera II
13 Maintenance and calibration
Summary of Contents for madsen itera ii
Page 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Page 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Page 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Page 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Page 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Page 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Page 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Page 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Page 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...