Otoacoustic emissions:
EN 60645-6, Type 2
Patient Safety:
•
EN 60601-1, Internally Powered, Type BF, IPXO
•
U2601-1; CAN/CSA-C22.2 NO 601.1-90,
•
IEC 60601-2-26
•
IEC 60601-2-40
EMC:
EN 60601-1-2
App. 3.4
Power supply and battery
Device
Supply voltage:
Nom. 3.70 V,
Max. 4.20 V,
Min. 3.20 V (measured with device load)
Maximum battery power con-
sumption:
1.5 W while measuring
Estimated battery life:
8 hours of continuous use (based on a typical use scenario. Actual use can influ-
ence the battery life time.)
Battery level indicator:
5-step battery level indicator
Battery
Caution
•
Use only rechargeable battery supplied by GN Otometrics A/S, Part no. 8-73-02400. Use of any other
battery may present a risk of fire or explosion.
Battery type:
Rechargeable Li-ion 3.7 V/1800 mAh (6.7 Wh), fully charged
App. 3.5
1077 docking station
The docking station is an optional accessory.
App. 3.5.1
Printer/modem interface
Interface type:
RS232
Connector type:
6-pol Mini Din
60
Otometrics - MADSEN Alpha OAE
App. 3 Technical Specifications
Summary of Contents for MADSEN Alpha OAE
Page 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Page 6: ...6 Otometrics MADSEN Alpha OAE ...
Page 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Page 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Page 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Page 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Page 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Page 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Page 68: ......