11.3
Calibrate Hardware Masking
Use a sound pressure-level meter to calibrate the masking level produced by the system when the system is installed and
as needed.
A. Set up the left transducer (insert earphone, headphone, or bone conductor) and the artificial ear (2cc or 6cc coupler
or artificial mastoid) with a Type 1 sound pressure-level meter according to the manufacturer's instructions.
B. Select Calibration >
Hardware Masking Calibration
.
C. Calibrate the hardware masking level.
Hardware Masking Calibration Dialog Box
1. Click and select the
Primary Trans-
ducer
(A).
2. Click
Up
or
Down
to adjust the sound
pressure- level meter to read 88 dB SPL
(+/-.5 dB).
3. Click
OK
.
D. Click
OK
to save the calibration.
Step 1asdfg
Step 2asdg
Step 3asdfg sdg
Step 4sdf gsd
Step 5fg sdfg
Step 6sdg
Step 7sdfg
Step 8sdfg
Step 9sdfg
Step 10sdfg
Step 11sdfg
Step 12sdfg
Step 13sdfg
Step 14sdfg
Step 15sdfg
Step 16sdf
Step 17g sdf
Otometrics - ICS Chartr EP 200
85
11 Diagnostic Calibrations
Summary of Contents for ICS Chartr EP 200
Page 1: ...ICS Chartr EP 200 Installation and Startup Guide Doc No 7 50 0570 EN 10 Part No 7 50 05700 EN ...
Page 4: ...Otometrics ICS Chartr EP 200 4 ...
Page 16: ...16 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Page 17: ...Otometrics ICS Chartr EP 200 17 3 Prepare Patient for Testing ...
Page 20: ...20 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Page 21: ...Otometrics ICS Chartr EP 200 21 3 Prepare Patient for Testing ...
Page 23: ...Otometrics ICS Chartr EP 200 23 3 Prepare Patient for Testing ...
Page 24: ...24 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Page 25: ...Otometrics ICS Chartr EP 200 25 3 Prepare Patient for Testing ...
Page 26: ...26 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Page 27: ...Otometrics ICS Chartr EP 200 27 3 Prepare Patient for Testing ...