Automatic activation of the thermal process for reducing bacterial count starts. Duration: 5 min.
•
When the thermal process is complete, the unit will switch to the irrigation mode (37° C).
•
When the irrigation mode is not used (handle is not removed from its support), the unit changes into the energy-sav-
ing mode after 5 min.
Fig. 6
Irrigation handle
A.
Lever for water flow
B.
Adjusting screw
4
Operation
Before switching on the unit, make sure that the handle is in its holder (jet must point to the rear of the unit).
4.1
Adjusting temperatures
•
Number of temperature levels: 3
–
One level fixed to 37° C irrigation temperature.
keys below the left and right switches (+)
–
Two variable temperature levels (20° C - 47° C) (48° C + 49° C only for testing purposes).
16
Otometrics - Aquamatic II
4 Operation