![otometrics AURICAL Aud Reference Manual Download Page 69](http://html.mh-extra.com/html/otometrics/aurical-aud/aurical-aud_reference-manual_1658652069.webp)
2.14.4.5
Minimum HTL per transducer type
The following table describes the minimum HTLs for which the Ambient Noise Assessor can measure ambient noise, for
various transducers. For stimuli below the given level, the Ambient Noise Assessor cannot distinguish between noise pro-
duced by the transducer and ambient noise. For stimuli below the minimum level, the ambient noise status indicator will
be yellow.
Transducer
Minimum HTL (dB HL)
Bone
30
Supra Aural, TDH 39/MX41
15
TDH39 with Peltor enclosure
0
Insert
-10
2.15
Test controls (keyboard, mouse)
2.15.1
PC keyboard controls
You can open an overview of PC keyboard shortcuts from the
OTOsuite Manuals
file.
OTOsuite Manuals
was installed on your PC when you installed OTOsuite. You can
open
OTOsuite Manuals
from the Windows
Start
menu.
Otometrics - AURICAL Aud
69
2 Navigating in the OTOsuite Audiometry Module