Test condition information
When conditions are good for data collection
TE TEST
will be shown at
the top of the screen. This fades in and out to show that a test is currently
running.
Noisy
appears if the noise level is above the noise reject level for a period
of time.
Check Probe Fit
is shown if the adjusted stimulus level falls outside the
accepted range.
The following table describes the test condition information which appear if
more than one condition is met:
Noise level indicator
The noise level indicator allows continuous monitoring of the noise level
during a test. The noise reject level is displayed numerically below the
indicator. Use the up/down arrow keys to adjust the noise reject level.
Cancel and End
Select
Cancel
at any time during a test to pause the test. This may be
useful if the ambient noise increases.
Cancel Test?
is displayed at the top
of the screen and three options are provided.
Yes
will terminate the test;
No
will continue the OAE recording and
Checkfit
will restart the test at the
Checkfit stage.
Select
End
at any time during a test to manually terminate the test.
Consistent
Stimulus
Highlighted
High Noise
Out of Range
Message
No
No
TE
TEST
No
Yes
Check
Probe
Fit
Yes
No
Noisy
Yes
Yes
Check
Probe
Fit
TEOAE test
Otoport OAE+ABR
User Manual for Advance and Screener