30.2
Confidence levels in TEOAE tests
In the case of TEOAE measurements, data points are aggregated in to
frequency bands. Normally this banding is performed as some proportion
of octaves. Because of this logarithmic banding of data there is a frequency
variance in the SNR required for any given level of confidence. This
leads to a more complex table for the SNR required for a single band.
Furthermore, this additional complexity is factored into the multiband tables
found in section
29.2.2
below.
30.2.1
Confidence for a single ½ octave band
Minimum SNR required for confidence Average of
level in a single band
all bands
Confidence
1kHz 1.5kHz 2kHz 3kHz 4kHz
99%
7dB 5dB 4dB 4dB 2dB 5dB
99.9%
11dB 8dB 6dB 6dB 4dB 8dB
99.99%
14dB 9dB 7dB 8dB 7dB 11dB
30.2.2
Confidence for multiple ½ octave bands
Most TEOAE test criteria test across all five ½ octave bands, between
1-4kHz:
Minimum SNR required for confidence level
Confidence 2 of 5 bands 3 of 5 bands 4 of 5 bands 5 of 5 bands
99% 2dB 0dB 0dB 0dB
99.9% 4dB 2dB 0dB 0dB
99.99% 8dB
4dB
2dB 0dB
Response detection
Otoport OAE+ABR
User Manual for Advance and Screener