CHAPTER TEN
DPOAE test
106
10.6.2
DP Growth graph
The DPG graph shows the DPOAE levels recorded at each of the
six intensity levels presented. The x-axis shows the target L2 levels
(45-70dBSPL). The y-axis shows the DPOAE level recorded in uPa or
dBSPL depending on the test setting (see chapter
17
Test set-up
). If the
uPa scale is used then the range is auto adjusted between 0-30 and 0-200
uPa depending on the largest DPOAE signal or noise level recorded.
If the line fit criteria are met then the best fit line is plotted in yellow on the
graph. If the uPa scale is selected then the DP threshold (Thr: ) is also
displayed.
All other features of the graph screen are as described in section
10.4.1
.
DPOAE test