
Otoport OAE+ABR
User Manual for Otoport DP+TE model
83
TEOAE test
Test Result
Description
TEOAE Pass
The data collected has met the criteria set. The
optimum test setting will depend on your application,
e.g. screening or clinical measurement
(see
15.2
Test settings
)
Note: The following results will only occur if a TEOAE Pass is not obtained,
providing test information feedback to the user
No Valid OAE
The data collected has not met the set pass criteria
and the test conditions were acceptable
Noisy
The noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit
Probe movement has been detected
Too Few Bands
Insufficient bands meet their pass criteria
Stopped Too
The test was ended manually before the
Soon
required minimum amount of data has been collected
9.4
TEOAE test results
The following table lists all possible test results with the associated result
graphic and gives an explanation of the circumstances under which each
result would be shown.
Select
Save
to skip to the
Patient Details
screen ready to
Save
.
Select
Review
to view the data of the completed test, prior to saving.
Select
Cancel
to discard the test result. A prompt requiring confirmation
will be shown before the result is discarded.