Mode of operation
26
Mode of operation
26.1
DPOAE test
Parameter
Description
Checkfit stimulus
80µs positive broadband square wave pulse with an intensity of
64dB pe (peak equivalent) in a 1cc cavity.
Sample rate
25.6kHz
Sample buffer
80ms (gives 2048 points)
FFT frequency
bin
25Hz
DP Noise calculation
DP noise from the five spectral points above and the five points
below the DP frequency. In the ‘Rapid’ noise mode, the mean of
these points gives the noise level. In the ‘Standard’ noise mode
the mean of these points plus two standard deviations gives the
noise level.
Noise rejection calculation
The noise level for noise reject is calculated from the difference
between consecutive 80ms samples.
Frequency ratio
F2= 1.22 F1