Ostec 4434-801-59962935-08 Operating Manual Download Page 16

 
 
 
 
 
 
 
 
 

12 

 

Spectrum identification 

ZaIR 3.5

 allows for identifying compounds by their IR spectra, name or other data by 

searching the spectrum libraries (libraries are not part of the system and are supplied 

separately). Click on the 

Spectrum Search

  button or press 

F5

  to start searching an open 

spectrum. In case of multiple open spectra, the first one is used. The search time depends on 

the number of search libraries and computer capacity and does not usually exceed two 

seconds. The button remains clicked during the search. If you unclick it before the search 

finishes, the search will be suspended at the current phase. 

The search results are displayed in the list on the left with the higher degree of similarity 

to the searched spectrum at the top {the similarity degree is displayed in the 

Rate

 column). 

While viewing the search results, the found spectra open as a second {red) curve. 

Instead of search by a full spectrum, you can search by a part of it using the 

Search 

bandwidth

 button. Clicking on it covers the spectral image with dark "curtains" that hide the 

out of search spectrum sections. You can pick and move the curtains by dragging their edges. 

Additional parameters are displayed in the 

Search Settings

 window that pops up after 

you click the button with the same name. The 

Spectrum

 tab shows the maximum number of 

the recordings to be selected and the minimum percentage of match rate. Peak positions and 

heights to be considered are also set as a percentage of the maximum possible values. The 

initial values are 85 and 25%, i.e. the peak positions are considered to a significantly greater 

degree than their height, which helps identify spectra recorded under different conditions. 

The 

Acceleration

 option involves all the CPU cores and multiplies the search speed almost by 

the number of the cores. The 

Reset to initial values

  button restores the default values 

applicable to ordinary tasks. 

The other parameters under the tab refer to searching by mixtures. The command itself 

is in the menu that drops down on clicking the side arrow to the right of the search button (in 

other words, the button menu). Searching mixtures involves the ordinary search by spectrum. 

The first entries of the search response, if their match rate is less than the preset value, are 

searched again after substracting the spectrum from the query spectrum. As a rule, the preset 

number of entries for a mixture search is smaller than that for an ordinary search (3–7) and 

the rate is bigger {50–80%). 

Note that a correct subtraction requires that the query spectrum  be within the 

absorption spectrum and reduced to it if it was in a different scale. Additionally, the spectrum 

sections outside the absorption bands should be close to zero, otherwise the base line needs 

adjustment. If it is not done manually, you should turn on the 

Subtract base line

 option. Note 

that after an adjustment (including automatic) the spectrum is deemed changed and you will 

be prompted to save the spectrum when closing it. 

Turn on the 

Only mixtures

 option only in case you are sure that your sample is not a 

pure substance. To eliminate results with single dominating substances, you can set the 

minor component minimum content (the 

Component

 parameter). 

Fig. 5 shows an example of mixture search for a mixture of two drugs. The search results 

start with the spectra that are a linear combination of the library curves. Such an "artificial" 

spectrum is marked with a"+" sign in front of the match rate. Click the sign to open the current 

line for the other lines to display links to its substance components (the list on the left).

 

 

Summary of Contents for 4434-801-59962935-08

Page 1: ...IROS 05 IR Fourier Spectrometer Operating Manual FSVE 443430 005RE 2020 Ostec Instruments Moscow...

Page 2: ...specialists who have read this manual are qualified to work on electrical devices up to 1 000 V and briefed on workplace safety rules The spectrometer involves operations with hazardous 220 V 50 Hz c...

Page 3: ...n 6 Device software installation 6 First run 7 Spectrum recording 7 Spectrum reading and copying 9 Spectrum identification 12 Viewing spectra 15 Spectrum editing 18 Calibration 19 Creating and editing...

Page 4: ......

Page 5: ...434 801 59962935 08 Please see spectrometer operating conditions in Table 1 Spectrometer s electromagnetic interference resistance complies with GOST 30804 6 1 2013 and radiated interference with GOST...

Page 6: ...ise ratio RMS in standard conditions in1 min at a resolution of 4 cm 1 in full bandwidth of 2000 2200 cm 1 40 000 Signal gain 1 2 4 8 Resolution 0 5 1 2 4 8 cm 1 Scanning speed spectra per minute For...

Page 7: ...1 Spectrometer optical scheme The inlet optical system 2 converts the IR source 1 radiant energy into a convergent annular beam and makes it focus on the inlet port of the interferometer 3 consisting...

Page 8: ...log to digital conversion is done at equal intervals of the optical path difference intervals The path difference reference scale is generated in a reference channel that consists of a HeNe laser the...

Page 9: ...t 2 the Scan light 3 peripheral equipment connection ports 4 sample compartment 5 MCT detector compartment cover 6 spectrum recording control buttons Note The photo displays the device in basic config...

Page 10: ...right sides The Power switch on the rear side of the device case near the 220 V power connector Device software installation First turn on the computer Run setup exe on the USB flash drive that comes...

Page 11: ...r icons there indicate scan errors Single errors can be caused by the device being pushed or shaken If errors keep occurring or the LED on the case stops blinking and the software window indicator loo...

Page 12: ...the beams reflected from an empty cell distort the spectrum Then click the Background Spectrum button or press Shift F9 The spectrum recording time is proportionate to the number of the scans and can...

Page 13: ...triangle to the right of the button and select you add on from the list Note that the scan number resolution and other recording settings apply to the current selection so switching to another add on...

Page 14: ...You can make any folder a home folder by opening it a clicking a context menu command Thus you can navigate the folder system on the disk through the list You can also open logical spectrum folders th...

Page 15: ...ial if you edit it the file is renamed As the Number field has the line number entering a number in this field results in the line under this number selected e g you can use this feature to get a spec...

Page 16: ...The Acceleration option involves all the CPU cores and multiplies the search speed almost by the number of the cores The Reset to initial values button restores the default values applicable to ordin...

Page 17: ...cating the number of each of them or the range using a hyphen e g C17 N3 5 02 Cl note that you do not have to use 1 Add to your formula if you want it to include any other elements except the listed o...

Page 18: ...ick the View button you can display them in advance Use the Compare button for comparison with the current spectrum and band coloring Use the Purge button to completely delete a peak list The From cur...

Page 19: ...rve The last open spectrum is marked as number 1 the previously opened spectra are numbered in an ascending order by adding one Any curve can become number one if you click on it You can also change t...

Page 20: ...ht Note that these variants do not affect the view if only one curve is open Sample spectra can be displayed both in the transmittance scale the quotient of two spectra current and background for each...

Page 21: ...he dependence of the spectrum current ordinate on the abscissa and vice versa Current X and Y are marked with dashed lines Their values are specified in the text fields at the graph edges You can use...

Page 22: ...it will change the base line or amplitude of the current spectrum in the area of the cursor moving If you magnify the curve images then your edits will be more accurate Plotting border lines involves...

Page 23: ...luted multiple times and recorded again Continue diluting until the peak of the substance to be identify appears on the spectrum For greater accuracy we recommend that you repeat the entire procedure...

Page 24: ...calculating concentration for that peak Consequently if a peak has graph a relevant concentration will show up near the peak height Creating and editing libraries As noted above spectrum libraries are...

Page 25: ...rogram If it does not help unplug and plug the USB cable back in Next turn off the device wait for a minute turn it on again and restart the program If it does not help restart the computer 2 Fault sc...

Page 26: ...rding settings set a different number where necessary If the channel number is correct check the sample compartment for foreign items or pollution 6 Fault software failure Manifestation the program do...

Page 27: ...ery two years Use the Calibration command in the Recording menu or the toolbar button The procedure is performed in the Master mode and consists of five steps Note that you may take measurements in an...

Page 28: ...equires a reference film and a comparison spectrum If you do not have them contact the spectrometer manufacturer or get a thin polystyrene layer ca 0 25 mm and record its spectrum and set the spectrum...

Page 29: ...n measuring precision are assessed at the next step This measurement is the most time consuming and can take a few minutes Its results are entered in the parameter list At the final step click the Rea...

Page 30: ...onditions are the same as the operating conditions Table 1 Place the spectrometer at least 10 cm away from the warehouse floor and walls and at least 50 cm from the heat sources Transportation and sto...

Page 31: ...measuring attenuated total reflection and specular diffuse reflection spectra It consists of a hold down 1 FTIR and SDR flange seats the seat shown covered with a plug flange 2 in Figure 12 The plug...

Page 32: ...remove the sample and clean the hold down tip and crystal surface with an alcohol or other solvent wetted cotton pad Liquid samples are placed at the FTIR element center in the form of a tiny drop do...

Page 33: ......

Reviews: