THOMAS SPLINT CLEANING INSTRUCTIONS

Cleaning can be achieved by either manual or automated methods - manual cleaning should only be undertaken 
when automated methods are inappropriate or unavailable. Automated cleaning is a controlled process that will 
allow more consistent results. Where available the use of mechanical cleaners such as washer-disinfectors and 
ultrasonic tanks is preferred to the manual cleaning of items. The cleaning method should be assessed to ensure 
the effectiveness of the cleaning process without damage to equipment.

Medical Devices Agency guidelines on the method of decontamination taking into account infection risk from the 
reprocessed device to the patient

Risk 

Application of an item 

Recommendation

High 

In close contact with a break in the skin or mucous membrane. Introduced into 
sterile body areas. 

Cleaning and Sterilisation

Medium 

In contact with mucous membranes. Contaminated with particularly virulent or 
readily transmissable organisms prior to use on immunocompromised patients.

Sterilisation or disinfection required. Cleaning 
may be acceptable in some agreed situations

Low

In contact with healthy skin. Not in contact with patient. 

Cleaning (physically) removes infectious 
agents without necessarily destroying them.

CLEANING GUIDELINES

Please refer to guidelines below for washing instructions. 

MECHANICAL CLEANING -THERMAL WASHER DISINFECTORS (REPEATED EXPOSURE TO WET HEAT)

Initial Clean - at or below 35°C. Followed by a hot water disinfection rinse where the surface temperature of the 
item processed should reach a minimum temperature of 71°C for a minimum of 3 minutes, 80°C for 1 minute or 
90°C for 1 second (HC(91)33 and BS 2745)

MANUAL CLEANING - NON IMMERSION 

A warm water/detergent solution at correct dilution
A clean, disposable, absorbent, non-shedding cloth or mechanical drying facility (e.g. drying cabinet or industrial 
hot air dryer)
A chemical neutraliser, first aid kit and eye wash bottle, in case of splashing with detergent.

STEAM STERILISATION 

Direct contact between the material being sterilised and 
pure dry saturated steam at the required temperature in 
the absence of air. The recommended combinations of 
time and temperature are given in HTM 2010, Part 3:

Frame Measurements for Steam Sterilisation  
(with side members fully retracted)  

 
 
NOTE: 

The Thomas Splint System is designed for long  

term use, however regular inspection of adjustment areas  
and frame body is advised. Please replace if necessary.

THOMAS SPLINT SLINGS

All Thomas Splint slings are single patient use and must be discarded  
at the end of the treatment.

Thomas Splint Component

Recommended cleaning method

Thomas Splint side members

Cleaning & Steam Sterilisation

Thomas Splint Thigh Hoops

Thermal Washer Disinfector

Thomas Splint Diposable Slings and Hoop Covers

Single Patient Use 400 wash

Sterilisation Temperature (°C)

115

121

126

134

Max Allowance Temperature (°C)

118

124

129

137

Minimum Holding Time (minutes)

30

15

10

3

Length

Width

Adult

80

21 cm

Paediatric

56

19cm

40C

 Wash 40°C 

Do Not Blech 

Dry Flat, Do Not Wring

FOR SINGLE

PATIENT USE

 Do Not Iron 

Single Patient Use Only

Össur Americas

Össur Nordic

Össur Europe BV – Italy

27051 Towne Centre Drive

P.O. Box 67

Via Baroaldi, 29

Foothill Ranch, CA 92610, USA

751 03 Uppsala, Sweden

40054 Budrio, Italy

Tel: +1 (949) 382 3883

Tel: +46 1818 2200

Tel: +39 05169 20852

Tel: +1 800 233 6263

Fax: +46 1818 2218

Fax: +39 05169 22977

Fax: +1 800 831 3160

[email protected]

[email protected]

[email protected]

Össur Iberia S.L.U

Össur APAC

Össur Canada

Calle Caléndula, 93 -

2F, W16 B

2150 – 6900 Graybar Road

Miniparc III

No. 1801 Hongmei Road

Richmond, BC

Edificio E, Despacho M18

200233, Shanghai, China

V6W OA5 , Canada

28109 El Soto de la Moraleja,

Tel: +86 21 6127 1707

Tel: +1 604 241 8152

Alcobendas

Fax: +86 21 6127 1799

Fax: +1 866 441 3880

Madrid – España

[email protected]

Tel: 00 800 3539 3668

Össur Europe BV

Fax: 00 800 3539 3299

Össur Australia

De Schakel 70

[email protected]

26 Ross Street,

5651 GH Eindhoven

[email protected]

North Parramatta

The Netherlands

NSW 2151 Australia

Tel: +800 3539 3668

Össur UK Ltd

Tel: +61 2 88382800

Tel: +31 499 462840

Unit No 1

Fax: +61 2 96305310

Fax: +31 499 462841

S:Park

[email protected]

[email protected]

Hamilton Road

Stockport SK1 2AE, UK

Össur Deutschland GmbH

Tel: +44 (0) 8450 065 065

Augustinusstrasse 11A

Fax: +44 (0)161 475 6321

50226 Frechen

[email protected]

Deutschland

Tel: +49 (0) 2234 6039 102

Fax. +49 (0) 2234 6039 101

[email protected]

Össur hf.

Grjótháls 1-5
110 Reykjavík
Iceland

WWW.OSSUR.COM

©Copyright Össur 2017

Rev. 1

IFU 1322_001

THOMAS SPLINT

Instructions for Use

Reviews: