![OrthoAccel Technologies AcceleDent Aura Directions For Use Manual Download Page 22](http://html1.mh-extra.com/html/orthoaccel-technologies/acceledent-aura/acceledent-aura_directions-for-use-manual_1654591022.webp)
22
Direction for Use AcceleDent® Aura
Technical Description
The following is a technical description for the AcceleDent Aura. This
information is intended to provide environmental operating conditions,
transport/storage conditions, as well as IEC 60601 equipment ratings,
and electrical safety classification for the included equipment. Conditions
should fall between the ranges detailed below.
Warning: No modification of this equipment is allowed.
Environmental operating conditions:
Ambient temperature range 5° C - 40° C
Maximum temperature of any external component at
maximum ambient temperature: 46°C
Relative humidity range 15% - 93%
Atmospheric pressure range 700 hPa – 1060 hPa
Transport and Storage Conditions:
Minimum Ambient Temperature: - 20° C
Maximum Ambient Temperature: 60° C
Maximum Humidity: 93%
Type BF Applied Part
The power adapter technical information is detailed below.
Part Number: PSA05F-050Q
Output Voltage: 5V
Current Limit: 1 A
AC Input Voltage Range: 100-240 V
AC Input Frequency: 50 - 60 Hz
Protection Class: Class II Equipment
Activator “ON/OFF” button is a push-push button to turn
activator on and off.